This report is being supplemented to provide additional information based on the device evaluation, the legal manufacturer's final investigation updated fields: d8, d9 date returned to mfg., h2, h3,h6 (type of investigation, investigation findings and investigation conclusions) and h11.The device was returned to olympus for inspection and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: damage or deterioration of parts due to wear and tear, without any design or manufacturing issue and component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|