This regulatory report is being submitted due to retrospective review through capa 624392.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Functionally, the battery was tested using the test handle.The test handle would power on and the display and was blue with squares across the screen, however an image was not visible on the screen.The led did illuminate from the camera.It was reported that the device had no display.The reported issue was confirmed.The most likely cause was traced to component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|