• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIRCRAFT MEDICAL LIMITED MCGRATH; LARYNGOSCOPE, RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIRCRAFT MEDICAL LIMITED MCGRATH; LARYNGOSCOPE, RIGID Back to Search Results
Model Number 340-000-000
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/07/2024
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter the video laryngoscope's battery had 119 minutes left, but no display was projecting before use.There was no patient involvement.
 
Additional Manufacturer Narrative
This regulatory report is being submitted due to retrospective review through capa: 624392.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Functional testing found that the battery was inspected, and there were no visual irregularities with the battery.Next, a 10 second battery test box (cl-15077) was used to test the battery.First, the initial voltage was found, which was 3.67 v.Then, a 10 second battery test was conducted.The battery was found to be a bad battery based on this test.The customer¿s battery was then put into the ao3 test mcgrath (cl-16435).The display did not work, though the led did work without any issue.The customer¿s battery was also put into the ao2 test mcgrath (cl-16058), the battery did not power did power on the device.However, the display was dim and wavy.Though, the led did work without any issue.It was reported that the device display was blank.The reported issue was confirmed.The most likely cause was traced to component failure.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCGRATH
Common Device Name
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
AIRCRAFT MEDICAL LIMITED
7 cross way
dalgety bay,fi KY11 9JE
UK  KY11 9JE
Manufacturer (Section G)
AIRCRAFT MEDICAL LIMITED
7 cross way
dalgety bay,fi KY11 9JE
UK   KY11 9JE
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key21462550
Report Number3010244187-2025-00378
Device Sequence Number8189393
Product Code CCW
UDI-Device Identifier15060272980068
UDI-Public15060272980068
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/21/2022
Device Model Number340-000-000
Device Catalogue Number340-000-000
Device Lot NumberH20112004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/08/2024
Initial Report FDA Received Date02/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/21/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-