| Model Number |
P10P-001 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Hemolysis (1886)
|
| Date of Event |
01/27/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The excor blood pump pu valves; 10 ml in/out; ø 6 mm; sn (b)(6) was in use on the patient from (b)(6) 2025 to date.The event occurred on (b)(6) 2025, which is (14 days) after the pump was placed on the patient.We have reviewed the production records of the excor blood pump and the pump was produced according to our specification.
|
| |
|
Event or Problem Description
|
|
The site contacted berlin heart inc.(bhi) clinical affairs (ca) on 2025-01-28 to report that the patient being supported with an excor pediatric vad system continued to have hemolysis since the implant on (b)(6) 2025.The ldh value showed >1000 u/l, which is 2.5 times higher than upper limits of normal and plasma free hemoglobin value was also noted to be 2.5 times higher than upper limits of normal.According to the site, the excor blood pump pu valves; 10 ml in/out; ø 6 mm; sn (b)(6) performed as intended with complete fill and ejection.There were no abnormal sounds from the excor blood pump.
|
| |
|