| Catalog Number |
G136 |
| Medical Device Problem Codes |
Restricted Flow rate (1248); Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
While using a cavitron plus g136, they allege that they have low water flow and the handpiece is heating up, no injury resulted.
|
| |
|
Additional Manufacturer Narrative
|
|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Additional Manufacturer Narrative
|
|
03/18/25 repair tech: tl.W4h water sol,iso valve damaged.Poor water pressure regulation, debris buildup in the water filter.Debris buildup in the handpiece cable, corroded contact pins on the steri-mate handpiece.Replaced damaged/worn components and recalibrated unit to factory specs.Qa -ds.Handpiece # :old 201404 -- new 202407.Handpiece cable # :old 05220 -- new12240.
|
| |
|