| Catalog Number |
G310SCLR |
| Medical Device Problem Code |
No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
While using a cavitron 300 series g310 they allege that they have no water, the unit is getting hot and the handpiece is getting hot; no injury resulted.
|
| |
|
Additional Manufacturer Narrative
|
|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(6) 2025 (b)(6).Hpc# (b)(6).Sterimate# (b)(6).E32 softward, e1code upgrade/update.The software probably cause confusion of operation.Updated the software to the current version.Tested and recalibrated to manufacturing specifications.(b)(6).
|
| |
|