• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENTII TRITANIUM MULTI 56F; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENTII TRITANIUM MULTI 56F; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 709-04-56F
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Ambulation or Postural Difficulties (2544)
Date of Event 01/30/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Reported event: an event regarding disassociation involving a trident shell was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Device history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: it was reported that the patient was revised due to disassociation of the mdm liner from the shell.The exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Event or Problem Description
Revised a right hip liner and head due to the mdm metal liner disengaging from the shell and overall instability.He tightened one of the acetabular screws a couple of turns and re-implanted a new mdm liner and went from a -4mm head to a +12mm head.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENTII TRITANIUM MULTI 56F
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
juana hiciano
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key21464480
Report Number0002249697-2025-00235
Device Sequence Number8822196
Product Code LPH
UDI-Device Identifier07613327343977
UDI-Public07613327343977
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K171768
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number709-04-56F
Device Lot Number89461301A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/30/2025
Initial Report FDA Received Date02/25/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age65 YR
Patient SexFemale
-
-