| Model Number |
DSX400T11C |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap/ bipap, and mechanical ventilator devices.The manufacturer received information alleging visualization of particles.No other clinical information or medical interventions were reported.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap/ bipap, and mechanical ventilator devices.The manufacturer received information alleging visualization of particles.The device was returned to the manufacturer's service center for further evaluation.During the evaluation of the device at the manufacturer service centre, the device was visually inspected and found no evidence of foam degradation.The device's downloaded logs were reviewed by the manufacturer service centre.There was no error code found.The manufacturer service center concludes that they couldn't confirm the customer's allegation and there was no visible foam degradation.Box h: evaluation method code, evaluation results code, component code grid and conclusion code grid has been updated.
|
| |
|