• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASAHI INTECC CO., LTD. ASAHI GLADIUS MONGO14; PTCA GUIDE WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASAHI INTECC CO., LTD. ASAHI GLADIUS MONGO14; PTCA GUIDE WIRE Back to Search Results
Model Number AP14R323S
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 02/11/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site: asahi intecc (thailand) co., ltd., pathum thani, thailand, registration number: 3003780911.Device evaluation could not be performed because the affected device was not returned.Lot history review revealed no anomaly relating to the reported event.No other similar product experience report was received from this lot.Based on the obtained information, lot history review, and referring to known similar events, it was presumed that tension exceeding the product design limit might have been applied on the subject gladius mongo14 guide wire during withdrawal attempts from the concomitant corsair pro xs microcatheter while the distal segment of the guide wire had been caught by the cto.Consequently, the guide wire was fractured.Although it was concluded that the reported event was not attribute to the product quality, it was concluded that stenting the wire fragments was an additional treatment.Capa: no capa will be taken.The instructions for use (ifu) states: [warnings].Observe movement of this guide wire in the vessels.Before this guide wire is moved or torqued, the tip movement should be examined and monitored under fluoroscopy.Do not move or torque the guide wire without observing corresponding movement of the tip; otherwise, the guide wire may be damaged and/or trauma may occur.In addition, ensure that the distal tip of this guide wire and its location in the vessel are visible during manipulations of the guide wire.Never push, auger, withdraw, or torque this guide wire that meets resistance.Torquing or pushing this guide wire against resistance may cause damage and/or tip separation of this guide wire or direct damage to a vessel.Resistance may be felt and/or observed under fluoroscopy by noting any buckling of the guide wire.If the prolapse of the guide wire tip is observed, do not allow the tip to remain in a prolapsed position; otherwise damage to the guide wire may occur.Determine the cause of resistance under fluoroscopy and take any necessary remedial action.If resistance is felt due to spasm, bending of the guide wire, or due to trap while operating this guide wire in the blood vessel or removing it, do not torque and/or pull the guide wire itself.Stop the procedure.Determine the cause of resistance under fluoroscopy and take appropriate remedial action.If the guide wire is moved excessively, it may break or become damaged, which may cause blood vessel injury or result in fragments being left inside the vessel.[malfunction and adverse effects] separation of the guide wire.
 
Event or Problem Description
It was reported that an asahi gladius mongo14 guide wire was used with an asahi corsair pro xs microcatheter to approach a moderately calcified and moderately tortuous chronic total occlusion (cto) caused by in-stent restenosis in the mid right coronary artery (rca) during percutaneous coronary intervention (pci).The gladius mongo14 guide wire met resistance at the proximal cap of the cto, but was advanced in an attempt to cross the lesion.When withdrawal from the corsair pro xs microcatheter was attempted, the distal segment of the gladius mongo14 guide wire was fractured.It was informed that the detached fragment was stented to the vessel wall, and the patient was discharged in a good condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASAHI GLADIUS MONGO14
Common Device Name
PTCA GUIDE WIRE
Manufacturer (Section D)
ASAHI INTECC CO., LTD.
3-100 akatsuki-cho
seto, aichi 489-0 071
JA  489-0071
Manufacturer (Section G)
ASAHI INTECC CO., LTD.
3-100 akatsuki-cho
seto, aichi 489-0 071
JA   489-0071
Manufacturer Contact
yukimi nagumo
3-100 akatsuki-cho
seto, aichi 489-0-071
JA   489-0071
MDR Report Key21465766
Report Number3003775027-2025-00081
Device Sequence Number3653419
Product Code DQX
UDI-Device Identifier04547327123709
UDI-Public(01)04547327123709(17)270831(10)240906A261
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K180784
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAP14R323S
Device Lot Number240906A261
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/12/2025
Initial Report FDA Received Date02/25/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Disability;
Patient Age75 YR
Patient SexFemale
-
-