• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. DJO SURGICAL; EMPOWR BLADE STEM, STEM, LATERAL, SIZE 15

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. DJO SURGICAL; EMPOWR BLADE STEM, STEM, LATERAL, SIZE 15 Back to Search Results
Catalog Number 425-92-015
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 01/06/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
See d4 expiration date, d10, h4, and h11.The agent reported "(peri-prosthetic fracture)".The previous surgery and the surgery detailed in this event occurred 2 months apart.This evaluation is limited in scope as the items associated with this investigation were not returned to djo surgical - austin for examination.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported devices were defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.The root cause of this complaint was a revision surgery due to periprosthetic fracture.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, inadequate soft tissue support, patient activities or trauma.
 
Additional Manufacturer Narrative
A return material authorization (rma) was issued to the customer requesting to have the intuitive surgical, inc.(isi) device returned.The event investigation is in progress.
 
Event or Problem Description
Revision surgery - due to periprosthetic fracture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DJO SURGICAL
Common Device Name
EMPOWR BLADE STEM, STEM, LATERAL, SIZE 15
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
stacey bonnell
9800 metric blvd
austin, TX 78758
5128346313
MDR Report Key21470357
Report Number1644408-2025-00050
Device Sequence Number20997402
Product Code KWY
UDI-Device Identifier00190446759041
UDI-Public00190446759041
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K223779
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number425-92-015
Device Lot Number4395A1018
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/06/2025
Supplement Date Received by Manufacturer02/10/2025
Initial Report FDA Received Date02/26/2025
Supplement Report FDA Received Date02/26/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/19/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
400-03-362 LOT: 866B2228; 400-03-362 LOT: 866B2228; 941-01-36G LOT: 599Z1176
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexFemale
-
-