• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIRCRAFT MEDICAL LIMITED MCGRATH; LARYNGOSCOPE, RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIRCRAFT MEDICAL LIMITED MCGRATH; LARYNGOSCOPE, RIGID Back to Search Results
Model Number 340-000-000
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/26/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This regulatory report is being submitted due to retrospective review through capa 624392.D10 concomitant product: 340-000-000, mcgrath 3.6v battery 340-000-000 (lot#10551402); 340-000-000, mcgrath 3.6v battery 340-000-000 (lot#h21111004); 340-000-000, mcgrath 3.6v battery 340-000-000 (lot#h23010704b); 340-000-000, mcgrath 3.6v battery 340-000-000 (lot#h23020404).H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Analysis of the device noted that the battery was installed in a test mcgrath mac.The battery would not power on the unit unless firm pressure was applied to the battery, indicating a contact issue at the terminals.The positive terminal was adjusted to make contact with the mac device.After this the device was powered on, battery indicator showed 246 minutes of battery remaining.The device was left on for 10 minutes, the device functioned properly during this time.It was reported that prior to use, the following video laryngoscope batteries were used and no display on the device screen, and no led lights lit.The reported issue was confirmed.The most likely cause was traced to component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, prior to use, the following video laryngoscope batteries were used and no display on the device screen, a nd no led lights lit.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCGRATH
Common Device Name
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
AIRCRAFT MEDICAL LIMITED
7 cross way
dalgety bay,fi KY11 9JE
UK  KY11 9JE
Manufacturer (Section G)
AIRCRAFT MEDICAL LIMITED
7 cross way
dalgety bay,fi KY11 9JE
UK   KY11 9JE
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key21470699
Report Number3010244187-2025-00446
Device Sequence Number11635761
Product Code CCW
UDI-Device Identifier15060272980068
UDI-Public15060272980068
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number340-000-000
Device Catalogue Number340-000-000
Device Lot NumberH21030704
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2024
Initial Report FDA Received Date02/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-