• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. ANGLED REAMER, DRIVE SHAFT; MANUAL INSTR, GENERAL SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. ANGLED REAMER, DRIVE SHAFT; MANUAL INSTR, GENERAL SURGICAL Back to Search Results
Model Number ANGLED REAMER, DRIVE SHAFT
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 02/18/2025, it was reported by a sales representative via (b)(4) that an ar-9597-10 angled reamer sleeve, 10° and an ar-9676 angled reamer, drive shaft fused together during surgery.This was discovered during a procedure on (b)(6) 2025, with no reported patient harm.Additional information has been requested.
 
Additional Manufacturer Narrative
Additional information: b5, d9, g3, h3, h6.The complaint allegation was confirmed.One unpackaged ar-9676 with an unknown batch was received for evaluation.The device arrived for evaluation, stuck inside the ar-9597-10 batch 3762223.The visual evaluation revealed nicks on the hudson connector.Functional testing cannot be performed as the device has the ar-9676 stuck inside the ar-9597-10 batch 3762223.The most likely cause for the reported failure can be attributed to misuse due to a force perpendicular to the drive shaft being applied while the device is spinning.This applied force creates an excess of friction and heat that ultimately has the potential to cause the device to seize from functioning as intended.Application of this force perpendicular to the drive shaft is not needed for the device to function, nor is it recommended that the user do this during normal clinical use.When used normally, this issue is unlikely to present itself, which suggests that the handling of the device has a large impact on whether or not it will seize.For this reason, the direct cause is considered to be misuse of the device.
 
Event or Problem Description
Additional information was received on 03/31/2025: this was discovered during a reverse total shoulder procedure on (b)(6) 2025.They opened a backup tray with ar-9597-10 and ar-9672 to complete the case.The case was delayed by one minute, with minimal anesthesia administered to the patient.No adverse effects on the patient were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGLED REAMER, DRIVE SHAFT
Common Device Name
MANUAL INSTR, GENERAL SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key21470932
Report Number1220246-2025-00690
Device Sequence Number3567746
Product Code LXH
UDI-Device Identifier00888867324954
UDI-Public00888867324954
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberANGLED REAMER, DRIVE SHAFT
Device Catalogue NumberAR-9676
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2025
Supplement Date Received by Manufacturer06/10/2025
Initial Report FDA Received Date02/26/2025
Supplement Report FDA Received Date06/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-