| Model Number |
N/A |
| Medical Device Problem Code |
Material Frayed (1262)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported we had a guide wire unravel in a patient.The metal tip zig zag wire became stuck in the patient needed to be removed with force.Patient upset, required xray to confirm no wire retained.Picc insertion aborted and patient refused another attempt.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and will be evaluated.A supplemental will be submitted with evaluation results.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a malfunctioned guidewire was confirmed.The product returned for evaluation was a guidewire.The investigation findings are consistent with damage caused by retraction of the guidewire against the bevel of the introducer needle.The returned product sample was evaluated and the guidewire was confirmed to be broken which allowed the outer coil wire to become unraveled.Microscopic examination of the fracture sites revealed the following: ¿ narrowing of the wire cross-section near the fracture site, which is a characteristic feature of a strong pull on the wire ¿ the weld tip of the wire was missing and could not be accounted for during the evaluation biological material was also seen on the wire which may have contributed to the observed guidewire fracture as retraction of the wire together with the biological material could have caused the pieces to become stuck.Normal movement of the guidewire is away from the sharpened bevel and will not damage the wire.If the guidewire direction is reversed, the guidewire is then pulled against the sharpened edge of the needle bevel and can cause shearing damage of the wire and/or cause the wire to become stuck within the needle.A second guidewire was returned but was found to be unremarkable in any failure type.This complaint will be recorded for future trending and monitoring purposes.
|
| |
|
Search Alerts/Recalls
|