MAUDE Adverse Event Report: BIOSENSE WEBSTER INC QDOT MICRO¿ CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC QDOT MICRO¿ CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Catalog Number D139501
Medical Device Problem Codes Insufficient Cooling (1130); Device Contamination with Body Fluid (2317); Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/24/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient underwent a pulmonary vein isolation (pvi) ablation procedure with a qdot micro¿ catheter and temperature titration occurred frequently.The ablation was stopped due to the temperature rose.This occurred after approximately 80 ablations.Flushing was conducted, but the char was attached to the catheter and could not be removed by wiping it off by hand.The issue was resolved by replacing the qdot micro catheter.No patient consequences were reported.
 
Additional Manufacturer Narrative
On 27-feb-2025, bwi received additional information regarding the event.The irrigation issue was noted after pulmonary vein isolation ablation.Furthermore, on (b)(6) 2025, the bwi product analysis lab received the device for evaluation.The product investigation was subsequently completed.It was reported that a patient underwent a pulmonary vein isolation (pvi) ablation procedure with a qdot micro¿ catheter and temperature titration occurred frequently.The ablation was stopped due to the temperature rose.This occurred after approximately 80 ablations.Flushing was conducted, but the char was attached to the catheter and could not be removed by wiping it off by hand.The issue was resolved by replacing the qdot micro catheter.No patient consequences were reported.Device evaluation details: the device was returned to johnson & johnson medtech (j&j medtech) for evaluation.Following j&j medtech procedures, a visual inspection, irrigation and temperature and impedance test of the returned device were performed.Visual inspection revealed no damage or anomalies on the device.The device was connected to the generator and an ablation cycle was performed and the device was found working correctly.No temperature or impedance issues were observed.An irrigation test was performed, and the device was flushing correctly, no obstructed holes were observed.No irrigation issues were observed.A manufacturing record evaluation was performed for the finished device number 31448705l, and no internal action related to the reported complaint were identified.The high temperature, irrigation and char issues reported by the customer could not be replicated during the product investigation; other issues or circumstances may have occurred during the use of the device that may have affected its performance.The instructions for use (ifu) contain the following information: ¿when rf energy is interrupted for either a temperature or an impedance rise (the set limit is exceeded), the catheter should be removed, and the tip should be inspected for char/coagulum that may be present on the tip.If present, do not continue the procedure with the same catheter and replace the catheter.If no char/coagulum is present, flush the tip to ensure the irrigation holes are not plugged prior to reinsertion of the catheter inside the patient body.¿monitoring the temperature from the electrode during the application of rf energy ensures that the irrigation flow rate is being maintained.As part of johnson & johnson medtech's quality process, all devices are manufactured, inspected, and released to approved specifications.This product complaint will be addressed through the quality system.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QDOT MICRO¿ CATHETER
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key21471752
Report Number2029046-2025-00576
Device Sequence Number3652570
Product Code OAE
UDI-Device Identifier10846835016710
UDI-Public10846835016710
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P210027
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberD139501
Device Lot Number31448705L
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/29/2025
Supplement Date Received by Manufacturer02/28/2025
Initial Report FDA Received Date02/26/2025
Supplement Report FDA Received Date03/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/21/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-