1 of 3 reports.Other mfg report numbers: 3006697299-2025-00024, 3014334038-2025-00029.This report is for a cerelink icp probe 1l p bolt: id 826852: a facility reported a cerelink system started to have an unreliable increase in the pic value up to 40, patient had a ct scan exam which showed icp of 80.The real correct icp value for this patient was 15/20.From an angiographic exam, the catheter seemed bent.; therefore, it was removed and replaced on (b)(6) 2025; one day after implantation.Patient in stable condition after replacement.The monitor did not show any error message.
|
|
Updated fields: d4, d9, g3, g6, h2, h3, h6, h11.The cerelink sensor was not returned for evaluation and lot number information has not been provided; therefore, an evaluation of the device could not be performed, and device history record (dhr) could not be reviewed.The root cause(s) of the reported issue could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|