|
Zimvie complaint number (b)(4).The following fields have been updated: b2.Outcomes attributed to adverse event b4: date of this report b5: describe event or problem d4: additional device information g3: date received by manufacturer g6: type of report h1: type of reportable event h2: follow up type h3: device evaluated by manufacturer h4: device manufacturer date h6: adverse event problem h10: additional narrative zimvie received one (1) fnt510, (full osseotite tapered implant 5 x 10mm) for evaluation.Visual evaluation was performed.Implant fracture identified at the collar.Fractured unknown screw identified inside implant.Device history record (dhr) review was completed for the subject lot number 2019050683.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2019050683 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : functional : fracture : implant based on the investigation and risk management file review, the most likely root cause determined from the investigation was patient factors.Therefore, based on the available information, a device malfunction did occur.Implant fracture identified at the collar.Fractured unknown screw identified inside implant.The reported event was confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
|