| Model Number |
70XLTUP |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
Unintended Extubation (4564)
|
| Date of Event |
02/14/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant product (70xltcd, 70xltcd 7.0mm xlt trach cuff dist x1, lo# 202409031x).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was an air leakage in the cuff on both the 7.0mm trach cuff and the 7.0mm uncuff trach at a hospital.The trach were replaced every time it failed.Eventually, a different manufacturer tracheostomy was placed without incident.Tracheostomy tube remained in place until a planned trach change was done.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was an air leakage in the cuff of the two 7.0mm tracheostomy tube at a hospital.The trach were replaced every time it failed.Eventually, a different manufacturer's tracheostomy was placed without incident.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the returned unit was contaminated.An attempt was made to inflate the returned unit and failed.Further examination showed that there was a large tear on the main body of the cuff.The single wall thickness of the cuff was measured away from the damaged area and found to be within the blueprint specification.It was reported that there was an air leakage in the cuff of the two 7.0mm tracheostomy tube at a hospital.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: do not use any sharp instruments such as forceps or hemostats that would damage the cuff when tapering it.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|