SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD MAXPLUS PRESSURE RATED EXTENSION SET WITH NEEDLELESS CONNECTOR; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number MP5310-C
Medical Device Problem Codes Fluid/Blood Leak (1250); Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/22/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that bd maxplus pressure rated extension set with needleless connector was leaking.The following information was received by the initial reporter with the following verbatim: it was reported by customer that the second iv extension tubing did not stay attached to the iv hub.The primary rn reattached and changed iv dressing earlier.Later in the day she paged the iv team to place another iv due to leaking.When assessed, we found the piv had blood return and flushed freely and the iv tubing was not tightening done on the hub.
 
Additional Manufacturer Narrative
No product or photo was returned by the customer.The customer complaint of connection issues could not be verified due to the product not being returned for failure investigation.A device history record review could not be performed on material mp5310-c because the lot number is unknown.Due to no sample being received, an investigation could not be performed, and a root cause could not be determined.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.Your assistance in this matter has been helpful in trend identification and supporting our commitment to continuous quality improvement.
 
Additional Manufacturer Narrative
No product or photo was returned by the customer.The customer complaint of leakage could not be verified due to the product not being returned for failure investigation.A device history record review for material# mp5310-c and lot# 24129102 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 06dec2024.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.Due to no sample being received, an investigation could not be performed, and a root cause could not be determined.
 
Event or Problem Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAXPLUS PRESSURE RATED EXTENSION SET WITH NEEDLELESS CONNECTOR
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key21499403
Report Number9616066-2025-00392
Device Sequence Number7724077
Product Code FPA
UDI-Device Identifier10885403236686
UDI-Public(01)10885403236686
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K083472
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberMP5310-C
Device Lot NumberUNKNOWN
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/11/2025
Supplement Date Received by Manufacturer03/13/2025
Initial Report FDA Received Date02/28/2025
Supplement Report FDA Received Date03/13/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/06/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-