• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JABIL CIRCUIT SINGAPORE PTE LTD BIS; INDEX-GENERATING ELECTROENCEPHALOGRAPH SOFTWAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JABIL CIRCUIT SINGAPORE PTE LTD BIS; INDEX-GENERATING ELECTROENCEPHALOGRAPH SOFTWAR Back to Search Results
Model Number PMB4000
Medical Device Problem Codes Insufficient Heating (1287); Temperature Problem (3022)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/19/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that prior to use, a device became hot when plugged into the docking station, reaching temperatures over 117 degrees.The device and docking station were noted to become hot to the touch when charging up to 100%.Although the device exceeded temperature limits while charging, it cooled down upon reaching full charge.Troubleshooting steps included walking the customer and representative through the process, replacing docking station, and testing by a representative on-site.Despite these efforts, the device continued to overheat.There was no patient involved.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: b5, g3, h6 additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunctioned.Therefore, this event does not meet the reporting requirements in 21 cfr 803.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that prior to use, a device became hot when plugged into the docking station, reaching temperatures over 117 degrees.The device and docking station were noted to become hot to the touch when charging up to 100%.Although the device exceeded temperature limits while charging, it cooled down upon reaching full charge.Troubleshooting steps included walking the customer and representative through the process, replacing docking station, and testing by a representative on-site.Despite these efforts, the device continued to overheat.There was no patient involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIS
Common Device Name
INDEX-GENERATING ELECTROENCEPHALOGRAPH SOFTWAR
Manufacturer (Section D)
JABIL CIRCUIT SINGAPORE PTE LTD
16 tampines crescent
singapore 52860 4
SN  528604
Manufacturer (Section G)
JABIL CIRCUIT SINGAPORE PTE LTD
16 tampines crescent
singapore 52860 4
SN   528604
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key21501291
Report Number2936999-2025-00164
Device Sequence Number15458858
Product Code OLW
UDI-Device Identifier10884521808270
UDI-Public10884521808270
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 03/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPMB4000
Device Catalogue NumberPMB4000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2025
Supplement Date Received by Manufacturer03/04/2025
Initial Report FDA Received Date03/02/2025
Supplement Report FDA Received Date03/10/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-