HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; RETROVIEW VIDEO COLONOSCOPE Back to Search Results
Model Number EC-3490TLI
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/20/2025
Type of Reportable Event Malfunction
Event or Problem Description
The time of event is unknown.There was no report of patient harm.Air/water nozzle peeled off (black glue).
 
Additional Manufacturer Narrative
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the nozzle gluing missing.Based on the result, we concluded that it was caused due to the physical damage applied on the nozzle gluing.In addition, our technician confirmed that the bending rubber leak, and the bending rubber cut; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.As a result of confirming the detail as gfe, no response was obtained, so we cannot deny the possibility of the glue falls into the human body.Therefore, based on the technical report "hr-rpt-0585(nozzle)" and/or the risk analysis results, it was evaluated to submit mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Common Device Name
RETROVIEW VIDEO COLONOSCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
gurvinder nanda
3 paragon drive
montvale, NJ 07645
2015712318
MDR Report Key21501543
Report Number9610877-2025-51226
Device Sequence Number20838171
Product Code FDF
UDI-Device Identifier04961333166619
UDI-Public04961333166619
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K131855
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEC-3490TLI
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2025
Initial Report FDA Received Date03/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/30/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-