| Catalog Number |
US1415064 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure in the superficial femoral artery via contralateral approach, the pta balloon was allegedly ruptured.After that, expanded it with the shiden hp and completed the procedure.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.Per the reported event details, the target lesion was severely calcified.Therefore, it is possible the interaction between the lesion and the balloon contributed to the reported balloon rupture.However, the definitive root cause for the reported balloon rupture could not be determined based upon available information.Labeling review: the definitive root cause for the reported balloon rupture could not be determined based upon available information.H10: g3, h6 (method).H11: b5, d4 (unique identifier (udi) #).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure in a severely calcified superficial femoral artery via contralateral approach, the pta balloon allegedly ruptured when expanded below the nominal pressure.The procedure was completed using another device.There was no reported patient injury.
|
| |
|