| Model Number |
5330008 |
| Medical Device Problem Code |
Mechanics Altered (2984)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: product analysis of part # 5330008 : lot # ed22f012 visual and optical inspection confirmed the laser weld on one side of the handle has broken.This is causing the handle to shift putting pressure on the inserter shaft when inserted and removed.The damage to the inserter is consistent with overload.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Information was received from user facility via a manufacturer representative regarding an inserter.It was reported that during ste rile processing, an issue was identified with an inserter where the inner stylet could not be passed through the inserter handle.No patient was associated with this event.
|
| |
|
Search Alerts/Recalls
|