• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. TRAPEZOID TIBIAL TRAY SZ 4F/4T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. TRAPEZOID TIBIAL TRAY SZ 4F/4T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 204-04-44
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Osteolysis (2377)
Date of Event 08/14/2023
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, approximately 5 years and 3 months post the initial right total knee arthroplasty, the patient presented back to surgeon's office with complaints of their knee.Upon examination, poly wear and osteolysis was evident.The patient was scheduled for a revision right total knee arthroplasty, possible poly swap.The patient underwent a full knee revision to competitor's products.The poly showed early wear and the metal implants were loose on the bone.There was no reported breakage of a device.There was a greater than 45 minutes surgical delay, but there were no adverse events as a result.
 
Additional Manufacturer Narrative
The reason for the revision reported in case (b)(4) may be the result of prosthesis wear, femoral loosening, and tibial loosening, or due to inclusion of the polyethylene in the packaging recall.Failure of the cement used to secure the femoral component and the tibial construct to the respective bone, may have lead to loosening at the cement-implant interface; however, this cannot be confirmed.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and radiographs and relevant clinical information were not provided.D10: logic pts, size 4, 8mm (cat# 02-012-42-4008 / serial# (b)(6)), fluted stem extension 40l x 14 mm (cat# 204-34-04 / serial# (b)(6)), tibial augment block 1/2-sz 4 5mm (cat# 204-64-05 / serial# (b)(6)), tibial augment block 1/2-sz 4 5mm (cat# 204-64-05 / serial# (b)(6)), tibial stem ext.Screw (cat# 204-70-00 / serial# (b)(6)).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEZOID TIBIAL TRAY SZ 4F/4T
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key21503796
Report Number1038671-2025-01352
Device Sequence Number20834770
Product Code JWH
UDI-Device Identifier10885862047694
UDI-Public10885862047694
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K933610
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/14/2019
Device Catalogue Number204-04-44
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2023
Initial Report FDA Received Date03/03/2025
Date Device Manufactured04/15/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization;
Patient SexMale
-
-