• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. TRULIANT TIB FIT TRAY CEM SZ 3.5F / 2.5T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. TRULIANT TIB FIT TRAY CEM SZ 3.5F / 2.5T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-022-45-3525
Medical Device Problem Codes Unstable (1667); Naturally Worn (2988)
Health Effect - Clinical Codes Inflammation (1932); Pain (1994); Synovitis (2094); Ambulation or Postural Difficulties (2544); Joint Laxity (4526); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); Joint effusion (4849)
Date of Event 02/23/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10: (b)(6) - 02-022-51-3513 - truliant tib imp crc insert sz 3.5, 13mm; (b)(6) - 02-020-13-0235 - truliant cr cem fem cr cem left sz 3.5; (b)(6) - 200-02-35 - three peg patella 35mm.Multiple mdr reports were filed for this event, please see associated reports: 1038671-2024-01232.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, and femoral loosening.Or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported that approximately 63 months after a left total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, instability, pain, significant synovitis, femoral loose from the cement.Initial surgery and revision surgery operative notes were provided.Patient left the operating room in stable condition.No further issues or complications were reported.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIANT TIB FIT TRAY CEM SZ 3.5F / 2.5T
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
MDR Report Key21504079
Report Number1038671-2025-01355
Device Sequence Number14552956
Product Code JWH
UDI-Device Identifier10885862305848
UDI-Public10885862305848
Combination Product (Y/N)N
PMA/510(K) Number
K152170
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number02-022-45-3525
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/26/2024
Initial Report FDA Received Date03/03/2025
Date Device Manufactured06/28/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-