Model Number 9732
Medical Device Problem Code Missing Test Results (3267)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The patient/family (a1) was the initial reporter (e1), so personal information was not entered.No information was captured in section a4 as the customer's weight was not provided.The contour® next gen meter does not have an expiration date.Therefore, no information was captured in section d4 (expiration date).The warranty period of the meter is 5 years from the date of the original purchase.
 
Event or Problem Description
The customer contacted ascensia customer service to seek assistance with his contour® next gen meter.During the troubleshooting, it was found that one of the customer's blood glucose readings was not stored in the meter's memory.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.A replacement meter kit was sent to the customer.
 
Additional Manufacturer Narrative
Despite three follow-up attempts, the customer did not return the device for evaluation.Therefore, a device history record was reviewed for the suspected contour® next gen meter with serial # (b)(6), and no manufacturing anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR® NEXT GEN
Common Device Name
BLOOD GLUCOSE METER
Manufacturer (Section D)
ASCENSIA DIABETES CARE US INC.
100 summit lake drive
valhalla NY 10595
Manufacturer (Section G)
PT PHC INDONESIA (REGISTRATION NUMBER: 3009204541)
kawasan industri mm2100 blok o
cikarang barat bekasi
jawa barat, 17520
ID   17520
Manufacturer Contact
shweta gulati
100 summit lake drive
valhalla, NY 10595
9142361830
MDR Report Key21504292
Report Number1810909-2025-00032
Device Sequence Number8006742
Product Code NBW
UDI-Device Identifier00301939732018
UDI-Public(01)00301939732018(17)250531(10)DM03T347P
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K223293
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 04/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number9732
Device Catalogue Number90010386
Device Lot NumberDM03T347P
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/04/2025
Supplement Date Received by Manufacturer04/18/2025
Initial Report FDA Received Date03/03/2025
Supplement Report FDA Received Date04/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/02/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age73 YR
Patient SexMale
-
-