Model Number 328512
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Consumer reported, when she removes the needle shield, the needle comes off with it and gets stuck in the shield.Lot: 4114009, catalog: 328512, date of event: unknown, samples: yes cl.
 
Additional Manufacturer Narrative
Additional information was added to: b4, g6, h2, h3, h11.Correction to: h6 (component code, type of investigation, and investigation conclusions).Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 0.3ML 31GA 8MM 10BAG 500 WAL
Common Device Name
SYRINGE, PISTON
Manufacturer (Section D)
EMBECTA MEDICAL I LLC - HOLDREGE, NE
Manufacturer Contact
avital merl
300 kimball dr.
parsippany, NJ 07054
MDR Report Key21504479
Report Number3024508819-2025-00112
Device Sequence Number7901957
Product Code FMF
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K212499
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number328512
Device Lot Number4114009
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/05/2025
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date03/03/2025
Supplement Report FDA Received Date03/24/2025
09/07/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-