| Catalog Number |
050-87215A |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Peritonitis (2252)
|
| Date of Event |
02/12/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.Clinical review: there is a temporal relationship between peritoneal dialysis utilizing the liberty select cycler with liberty cycler set and the patient event of peritonitis.However, there is no documentation in the complaint file to show a causal relationship between the adverse event and the use of the liberty select cycler and cycler set.Additionally, there is no allegation or evidence of a device malfunction or deficiency, or cycler set defect reported for this event.Although no information was provided related to culture results or the determined cause of the reported peritonitis, all pd patients are at risk of peritonitis due to a compromised immune system and the increase of potential for microbial contamination from dialysis techniques.Most cases of peritonitis are caused by touch contamination by the patient with the pd catheter being a source of entry for organisms into the peritoneum.Based on the limited information and no allegation of a malfunction, deficiency, or defect the liberty select cycler and cycler set can be excluded as the cause of the patient¿s reported peritonitis event.
|
| |
|
Event or Problem Description
|
|
It was reported that this peritoneal dialysis (pd) patient was hospitalized on (b)(6) 2025 due to peritonitis.The patient underwent a pd catheter (not a fresenius product) removal.The patient was reported to be completing temporary in-center hemodialysis (hd).Additional information was obtained through follow-up with the pd clinic.The clinic had not received the patient¿s hospital records, and therefore did not have information pertaining to the dates of hospitalization, culture results, or antibiotic treatment.It was confirmed that the patient did have the pd catheter removed (unknown date) and that the patient is completing in-center hd.It was also confirmed that there were no reported cassette leaks or other issues with any fresenius products at the time of the peritonitis diagnosis.No further information was available.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the patient for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for all fresenius liberty cycler sets shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
|
| |
|
Event or Problem Description
|
|
It was reported that this peritoneal dialysis (pd) patient was hospitalized on (b)(6) 2025 due to peritonitis.The patient underwent a pd catheter (not a fresenius product) removal.The patient was reported to be completing temporary in-center hemodialysis (hd).Additional information was obtained through follow-up with the pd clinic.The clinic had not received the patient¿s hospital records, and therefore did not have information pertaining to the dates of hospitalization, culture results, or antibiotic treatment.It was confirmed that the patient did have the pd catheter removed (unknown date) and that the patient is completing in-center hd.It was also confirmed that there were no reported cassette leaks or other issues with any fresenius products at the time of the peritonitis diagnosis.No further information was available.
|
| |
|