| Catalog Number |
ECR45G |
| Medical Device Problem Code |
Failure to Form Staple (2579)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an unknown surgery, after fired, noted the staples malformed.Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 03/03/25.D4: batch #unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device ecr45g with lot number 314c51; and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 3/26/2025.The device/photo upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 04/22/2025 d4: batch #260c64.Investigation summary the product was returned for evaluation.Visual inspection was conducted on the returned reload.Visual analysis of the returned sample determined that one ecr45g reload was received with no apparent damage and fully fired.No functional test could be performed due to the condition of the reload.The event described could not be confirmed as the reload was received fully fired.Although no conclusion could be reach on the cause of the reported event, the instructions for use do contain the following caution: tissue thickness should be carefully evaluated before firing any stapler.Holding the jaws in place for 15 seconds after closing and prior to firing may result in better compression and staple formation.When positioning the stapler on the application site, ensure that no obstructions such as clips, stents, guide wires, etc.Are within the instrument jaws.A manufacturing record evaluation was performed for the finished device batch number 260c64, and no non-conformances were identified.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|