• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC BLUE HAND PIECE; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC BLUE HAND PIECE; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HPBLUE
Medical Device Problem Codes Failure to Read Input Signal (1581); Difficult or Delayed Activation (2577)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unknown procedure there was a hpblue failure.There were no patient consequences.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 3/3/2025.B3: event year only reported: 2024.D4 batch #: a9d537.Additional information was requested and the following was obtained: did the generator display any error messages and if so what were they? yes, they had an error message but don¿t know which one.Did the device pass the pre-test? yes.Had the device been working and then stopped? yes.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the device was returned with no apparent damage.It was connected to a generator, evaluated with a test instrument and the hand activation feature was non-functional.However, it worked properly with footswitch assembly.The instrument was disassembled to perform an internal electrical test that revealed a hand activation open circuit condition at the cable level, affecting the functionality of the handpiece.In addition, the moisture indicator was positive.The handpiece has a number of seals to prevent fluids from entering the housing."positive moisture indicator¿ describes a condition where water enters the handpiece cavity during the steam sterilization process.The primary path of ingress is the distal seal, this may be caused by a reduction of the compressive force on the distal seal.However, no definitive root cause could be drawn.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reach on the cause of the reported event.It is probable that the ingress of moisture affected handpiece functionality.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC BLUE HAND PIECE
Common Device Name
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key21506349
Report Number3005075853-2025-01675
Device Sequence Number4400992
Product Code LFL
UDI-Device Identifier20705036002151
UDI-Public(01)20705036002151
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K063192
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberHPBLUE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/28/2025
Initial Report FDA Received Date03/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/08/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATOR AND HAND PIECE.
Patient SexUnknown
-
-