D10: (2886092) 02-012-35-2009 - logic tibia ps mod insrt sz 2 9mm;p (4010919) 200-02-32 - three peg patella 32mm; (3933347) 02-012-41-2020 - logic tibia traptray cem sz 2f/2t.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
|
The revision reported may have been the result of prosthesis wear, femoral loosening, and tibial tray subsidence as stated in the operative notes.The prosthesis wear may have been due to malalignment between the implants, high contact stresses during knee flexion, third body wear, patient-related conditions, instability, or any combination of these possibilities.The aseptic (non-infected) femoral loosening may have been the result of an insufficient bond between the femoral component and the bone and/or patient-related conditions.The tibial tray subsidence may have been the result of poor bone quality, misalignment, a load transfer issue, or any combination of these possibilities.However, the reason for the reported prosthesis wear, femoral loosening, and tibial tray subsidence could not be confirmed as the devices were not returned for evaluation, and images/radiographs were unable to be obtained.H6: corrected investigation clinical codes.
|