MEDTRONIC NAVIGATION, INC NAVLOCK; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9733734
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/15/2024
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received information regarding a navigation system being used outside of a procedure.It was reported that the site had 2 instruments that were defective.The tracker was not seen by the positioner.The ratcheting straight handle had a worn handle. the cause of the damage was likely wear and tear from use.There was no patient involvement.
 
Additional Manufacturer Narrative
Product analysis # (b)(4) :2024-06-05 13:36:25 cdt webmremotews sfdcmnav product analysis task update tested by date: 2024-06-05 findings and conclusions: the ratchet mechanism is broken and has separated from the returned handle.As a result, the handle will not ratchet in either direction.Afc: nafc0118 - damaged tool; this regulatory report is being submitted due to retrospective review through capa 626390.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVLOCK
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key21507120
Report Number1723170-2025-01113
Device Sequence Number4500996
Product Code OLO
UDI-Device Identifier00763000054465
UDI-Public00763000054465
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K124004
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number9733734
Device Catalogue Number9733734
Device Lot Number190430
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/05/2024
Initial Report FDA Received Date03/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/30/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-