Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: investigation summary the product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device revealed that the edges of the tip of the protect sleeve 12/8 l188 had minor impacts, causing a deformation of its original shape.Additionally, scratches were observed on the external surface of the shaft due to normal use over approximately 12 years.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A dimensional inspection was not performed due to post manufacturing damage.A functional evaluation was not performed as it is not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the protect sleeve 12/8 l188 would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history part# 03.010.063 lot # 8690018 manufacturing site: werk hägendorf supplier; na release to warehouse date: 22 nov 2013 expiration date: na.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|