• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION OPTICROSS HD; CATHETER, ULTRASOUND, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION OPTICROSS HD; CATHETER, ULTRASOUND, INTRAVASCULAR Back to Search Results
Model Number H74939352040
Medical Device Problem Codes Material Twisted/Bent (2981); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/10/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that device entrapment occurred.The 99% stenosed target lesion was located in a severely tortuous and severely calcified mid circumflex artery.A 6f guidezilla ii, an opticross hd and a marvel guidewire were selected for use.During the procedure, it was noted that upon crossing the proximal lesion, the physician felt resistance, then turned off imaging and had to pull out all three devices as a unit.Furthermore, the marvel guidewire or guidezilla could not separate from the opticross.The procedure was completed with the use of a different non-boston scientific device.No patient complications were reported and the patient status was good post procedure.
 
Additional Manufacturer Narrative
G4: premarket / 510(k) #: k213593.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTICROSS HD
Common Device Name
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
7634941632
MDR Report Key21510677
Report Number2124215-2025-11550
Device Sequence Number8007503
Product Code OBJ
UDI-Device Identifier08714729960737
UDI-Public08714729960737
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K173284
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/11/2025
Device Model NumberH74939352040
Device Catalogue NumberH74939352040
Device Lot Number0033807882
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2025
Initial Report FDA Received Date03/03/2025
Date Device Manufactured04/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-