|
It was reported that a clearlink system, non-dehp continu-flo solution set leaked total parenteral nutrition (tpn) from the y-site (clearlink) at a "brisk rate".This was observed during patient use.The issue was further described as, "visible drip from 2nd y site down from the spike".New intravenous fluids were ordered to resolve the event.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Additional information: e4, g2 (add source), h3, h6, h10, h11.H11: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Pressure and clear passage testing was performed along with priming; no defects were observed that could generate the reported condition.Functional flow rate, gravity, flow rate using in-lab spectrum iq pump, and back pressure testing on the clearlinks were performed with no issues noted; the set met specifications.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
|