• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B.T.I. BIOTECHNOLOGY INSTITUTE S.L.; Implant, endosseous, root-form

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B.T.I. BIOTECHNOLOGY INSTITUTE S.L.; Implant, endosseous, root-form Back to Search Results
Model Number IIPECA3585
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 03/13/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The batch documentation is checked and no anomalies are detected.Any manufacturing fault is ruled out.After examination of the implant, it can be verified that the implant does not show signs of misuse or signs of malfunction that could have contributed to the event.
 
Event or Problem Description
The clinician indicates that she placed the implant on (b)(6) 2024 and 1 month later it was noted that the implant had not osseointegrated.Patient with bone quality iii.In the event the patient experienced: infection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
Implant, endosseous, root-form
Manufacturer (Section D)
B.T.I. BIOTECHNOLOGY INSTITUTE S.L.
leonardo da vinci street,14
minano menor, alava 01510
SP  01510
Manufacturer (Section G)
B.T.I. BIOTECHNOLOGY INSTITUTE S.L.
leonardo da vinci street,14
minano menor, alava 01510
SP   01510
Manufacturer Contact
paula sanchez
leonardo da vinci street,14
minano menor, alava 01510
SP   01510
MDR Report Key21512445
Report Number3004417597-2025-00043
Device Sequence Number14578091
Product Code DZE
UDI-Device Identifier08435389813187
UDI-Public08435389813187
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K211952
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Dentist
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberIIPECA3585
Device Lot NumberBC01924.A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/16/2024
Initial Report FDA Received Date03/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR