| Catalog Number |
254500060 |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was difficulty with pin puller while pulling pins.Pin puller was sticking according to pa who used it.The surgical delay is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: it was reported that there was difficulty with pin puller while pulling pins.Pin puller was sticking according to pa who used it.The surgical delay is unknown.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device does not found signs of jammed at the attune pin puller presents and overall worn appearance consistent with normal and constant usage for a long period of time.A functional testing was performed using a styrofoam block and pin samples.The attune pin puller was able to securely grab and the spring was fully functional and allowed the puller to clamp down on the pin for extraction from the hard styrofoam and then open to release the pin as intended.The reported condition was not able to be replicated.A dimensional inspection was not performed as it is not applicable to the complaint condition.The overall complaint was not confirmed as the observed condition of the attune pin puller would have not contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11 additional narrative: corrected: h3.
|
| |
|
Search Alerts/Recalls
|