| Catalog Number |
72201500 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during set up, the dyonics power mini was overheating.There was a back-up device available, and no delay was reported.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: h2: updated data in h6.H3, h6: the reported device was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.A functional evaluation revealed the mdu worked as intended.No issues were found.No overheating noted.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause could not be determined since the reported malfunction could not be duplicated during the investigation.Factors that can contribute to the reported event include failure of a concomitant device.Based on this investigation, the need for corrective action is not indicated.
|
| |
|
Search Alerts/Recalls
|