| Catalog Number |
735-02 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the temperature probe was not reading accurately and arctic sun device rewarming very slowly.Per review of wo on 03mar2025, device was used on patient and moved to another unit.
|
| |
|
Additional Manufacturer Narrative
|
|
The complete initial reporter phone number is (b)(6).The complete initial reporter facility name is leeds general infirmary receipt & distribution, investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The reported product number is sold ous.The udi number for this product is not available.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported issue was confirmed.The root cause identified is a failed temp cable.The device was evaluated upon receipt.Temperature cable not working.Replaced temp cable.It was reported that device rewarming very slowly, but machine rewarming above 30 c within 10 minutes as per verification check.The as5000 system was evaluated for functionality and results were recorded.An electrical safety test was performed; as5000 system passed all tests, is functioning properly and is ready for use.(b)(4) rev 25 was reviewed for this investigation.The reported event is addressed in step #ra15.Based on this review the risk is within the expected range.The complaint or reported issue was confirmed through other elements of the investigation to not be labeling or packaging related.A dhr review is not required as the lot/serial number is unknown.Correction: d, h.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the temperature probe was not reading accurately and arctic sun device rewarming very slowly.Per review of wo on 03mar2025, device was used on patient and moved to another unit.Per follow up information received via mail on 03mar2025, the device was not here yet.Per follow up information received via mail on 04mar2025, awaiting unit¿s arrival.The issue was not resolved yet.There was no reported patient impact the device was sent with reported fault of not regulating patient temperature.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: b, d, e, f, h.The initial reporter phone number is (b)(6).The complete initial reporter facility name is leeds general infirmary receipt & distribution.Investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the temperature probe was not reading accurately and arctic sun device rewarming very slowly.Per review of wo on 03mar2025, device was used on patient and moved to another unit.Per follow up information received via mail on 03mar2025, the device was not here yet.Per follow up information received via mail on 04mar2025, awaiting unit¿s arrival.The issue was not resolved yet.There was no reported patient impact the device was sent with reported fault of not regulating patient temperature.
|
| |
|