MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3543507 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Breast Discomfort/Pain (4504)
|
| Date of Event |
11/15/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: as of now there is no information regarding the explant or reoperation d4: udi: as the serial number for the device involved in the event was not provided, the full udi is currently not available.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a 63-year-old caucasian female patient underwent revision breast augmentation with 350cc mentor memorygel breast implant on the right side, and with 400cc mentor memorygel breast implant on the left side and experienced bilateral breast implant rupture, and bilateral breast pain postoperatively.Mri confirmed rupture.The patient has consulted with the healthcare professional.At the time of this report, mentor has received no information regarding explantation or an expected explantation date.This medwatch report is for the patient¿s right-sided device.
|
| |
|
Additional Manufacturer Narrative
|
|
On may 20, 2025, mentor became aware that the date of adverse event date was november 15, 2024, and date of surgery is (b)(6) 2025.Hence, following fields have been updated on this form: is required intervention has been selected under section b; field b2.Date of event under field b3 has been updated to november 15, 2024.Fields d6b for explantation date is (b)(6) 2025, field h6 health effect - impact code ¿device explantation¿ has been added, field h6 type of investigation has been updated to "device not returned." reason for device explant and/or reoperation: right pain and rupture.Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.This supplemental report is for the patient¿s right-sided device.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On july 29, 2025, device evaluation was completed as follows: mentor conducted a visual inspection of the returned device.Visual analysis of the returned device revealed that the breast implant was found to be ruptured, in addition, siltex cracking was noted on the edges of the rupture.The evaluation determined that the possible cause of the rupture is consistent with normal wear.Siltex cracking is defined as a material separation occurring in the siltex layer and can eventually progress through the shell of siltex devices.Siltex cracking is ultimately a result of high stresses.Most women undergoing augmentation or reconstruction with a mammary prosthesis will experience some pain postoperatively.While pain normally subsides in most women as they heal from surgery, it can become a chronic problem in other women.Chronic pain can be associated with a variety of factors.Surgeons should instruct their patients to inform them if there is significant pain or if pain persists.Pain is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria, and that normal wear may have caused the observed defect.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.This supplemental report is for the patient¿s right-sided device.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On july 2, 2025, the mentor failure analysis lab received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.This supplemental report is for the patient¿s right-sided device.Manufacturer¿s reference number: (b)(4).
|
| |
|