MAUDE Adverse Event Report: STRYKER-ENDOSCOPY LAKELAND 5300; SINGLE-USE REPROCESSED ULTRASONIC SURGICAL INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER-ENDOSCOPY LAKELAND 5300; SINGLE-USE REPROCESSED ULTRASONIC SURGICAL INSTRUMENTS Back to Search Results
Model Number HAR1120
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the blade broke while cutting liver tissue.As reported, the device was replaced with a stryker device.Per the sales representative, the blade was recovered in full, the procedure was not altered in any way, and no imaging was used post care given the blade was fully recovered.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
 
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Upon inspection of the received complaint device, the jaw blade was observed to be fractured.The fractured tip of the blade was returned with the complaint device.The device inspection and functional testing indicated that the complaint was confirmed for the reported failure mode(s).A review of the dhr supports that the device met all inspections and test criteria prior to release from stryker.The reported event could be attributed to: - shipping damage, handling damage.Jaws/blade subassembly damage or separation.Too much force or torque applied to instrument, or grasping/pulling.Scalpel blade tip or jaw broken or damaged, cleaning instrument or blade tip with abrasives.Applying pressure between instrument blade and tissue pad without having tissue between them.Keeping clamp arm open when backcutting or while blade is active without tissue between blade and tissue pad.Incidental and prolonged activation against solid surfaces, such as bone, metal or plastic.Attempting to bend, sharpen, or otherwise alter the shape of the blade.The instructions for use (ifu) state: verify compatibility with generators.Use device only with ethicon endo-surgery generator g11 (gen11) software version 2018-1 or later.Software revision can be found under ¿system information¿ in the generator g11 (gen11) ¿settings¿ menu.Refer to the generator g11 (gen11) operator¿s manual for more information.Avoid contact with any and all metal or plastic instruments or objects when the instrument is activated.Contact with staples, clips or other instruments while the instrument is activated may result in scratches on the blade, cracked or broken blades and premature blade failure.Audible high-pitched ringing, resonating from the blade, is an abnormal condition and an indicator that the blade is not operating properly.This may result in abnormally high shaft temperatures and user or patient injury.Do not attempt to bend, sharpen, or otherwise alter the shape of the blade.Doing so may cause blade failure and user or patient injury.Care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Clamping the tissue pad against the active blade without tissue on the full length of the blade will result in higher blade, clamp arm and distal shaft temperatures and can result in possible damage to the instrument.If this occurs, there may be an instrument failure, and the generator touchscreen displays a troubleshooting message.The reported event will continue to be monitored through post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
SINGLE-USE REPROCESSED ULTRASONIC SURGICAL INSTRUMENTS
Manufacturer (Section D)
STRYKER-ENDOSCOPY LAKELAND 5300
5300 region ct
lakeland FL 33815
Manufacturer (Section G)
STRYKER-ENDOSCOPY LAKELAND 5300
5300 region ct
lakeland FL 33815
Manufacturer Contact
andy dobos
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key21516172
Report Number0001056128-2025-00011
Device Sequence Number9560140
Product Code NLQ
UDI-Device Identifier07613327586565
UDI-Public07613327586565
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K233471
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHAR1120
Device Catalogue NumberHAR1120
Device Lot Number16463988
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/04/2025
Initial Report FDA Received Date03/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/10/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-