|
(b)(4) date sent 3/4/2025 d4 batch #112d84.Investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned reload.Visual analysis of the returned sample determined that the tcr75 reload was received along with its opened packaging with no apparent damage and with no instrument.The swing tab was found to be in the locked position and with 7 drivers up along the staple line.The reload was tested for functionality with a test device and achieved its firing sequence without any difficulties and it performed as intended.However, after the functional test, 1 staple was noted missing along the staple line, this staple do not create a fluid path.It is possible that improper handling of the reload caused the staples to fell out, the proper handling instructions should be used when removing and reloading cartridges into the device, please reference the instructions for use.It should be noted that 100% inspection is performed by means of an automated vision system to ensure staples presence; the swing tab is present and unlocked.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 112d84, and no non-conformances were identified.The lot#131d62 history records were reviewed and certified by external manufacturing that the manufacturing criteria were met prior to the release of the equipment.The certificate records are accessible through external manufacturing.
|