| Catalog Number |
620046160 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that plastic tabs were broken on a humeral head trial.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: d4: udi: as the catalog/model number was not provided, (01)gtin is not available.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d4 (catalog, lot, primary udi number), g4, h4.Corrected: d1, d2a, d3, g1.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary it was reported that plastic tabs were broken on humeral head trial.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found the tabs inside were broken.The broken fragments were not returned for evaluation.Damage observed can be consistent with the user applying excessive leverage/torque on the device in a circular pattern while disassembling the device.Inhance¿ shoulder system surgical technique states the following: "to remove the humeral head trial, assemble the head distractor to the impactor handle and place between the resection and the humeral head.To remove the offset taper adapter trial from the humeral head trial, rotate the offset taper adapter trial counterclockwise until the ramps unlock it from the head.From the e=eject position, additional counterclockwise rotation of the offset taper adapter trial will push it up the ramp and out of the humeral head trial.A functional test was not conducted due to not being applicable to the complaint condition a dimensional inspection was not conducted due to not being applicable to the complaint condition the overall complaint was confirmed as the observed condition of the humeral head trial 46x16 would have contributed to the complained device issue.Based on the investigation findings, the potential cause is traced to unintended use, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|
|