DEPUY IRELAND - 3015516266 ATTUNE SHIMS & GEN INST; ATTUNE INSTRUMENTS : INSTRUMENT CASES Back to Search Results
Catalog Number 254501704
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/13/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported by central sterile services department (cssd) that the plastic coating is delaminating from metal tray.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: according to the information received that "cssd is complaining the instrument tray as the plastic coating is delaminating from metal tray." the product was not returned to medtech orthopedics; however, photos were provided for review.See attachment "(b)(4) photos" the photographs attached were reviewed, however they do not represent the reported complaint.Therefore, the investigation could not draw any conclusions about the reported event due to the insufficient evidence provided.The device lot number is unknown; therefore, a device history review could not be performed.If the lot/serial number becomes available, the record will be reassessed.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the photographs provided are not representative of the reported complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot - the device lot number is unknown; therefore, a device history review could not be performed.If the lot/serial number becomes available, the record will be reassessed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary cssd is complaining the instrument tray as the plastic coating is delaminating from metal tray.See photos.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found the plastic coating covering the brackets, was peeling off.Additionally, the graphics on the inside and outside of the tray were pealing.Furthermore the lot number was unable to retrieved due to the worn condition.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.A manufacturing records evaluation (mre) was not performed as no lot number was retrieved for this device due to the condition (wear) of the returned device the overall complaint was confirmed as the observed condition of the attune shims & gen inst would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot a manufacturing records evaluation (mre) was not performed as no lot number was retrieved for this device due to the condition (wear) of the returned device.
 
Additional Manufacturer Narrative
Product complaint#: (b)(4).Investigation summary: according to the information received that "cssd is complaining the instrument tray as the plastic coating is delaminating from metal tray." the product was not returned to medtech orthopedics; however, photos were provided for review.The photographs attached were reviewed, however they do not represent the reported complaint.Therefore, the investigation could not draw any conclusions about the reported event due to the insufficient evidence provided.The device lot number is unknown; therefore, a device history review could not be performed.If the lot/serial number becomes available, the record will be reassessed.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the photographs provided are not representative of the reported complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the device lot number is unknown; therefore, a device history review could not be performed.If the lot/serial number becomes available, the record will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE SHIMS & GEN INST
Common Device Name
ATTUNE INSTRUMENTS : INSTRUMENT CASES
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key21518952
Report Number1818910-2025-03430
Device Sequence Number13190000
Product Code FSM
UDI-Device Identifier10603295423546
UDI-Public(01)10603295423546
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254501704
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/13/2025
Supplement Date Received by Manufacturer05/14/2025
05/20/2025
05/23/2025
06/12/2025
Initial Report FDA Received Date03/04/2025
Supplement Report FDA Received Date05/15/2025
05/21/2025
05/28/2025
06/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-