H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and will be evaluated.A supplemental will be submitted with evaluation results.
|
|
It was reported, had a picc placed on 23rd january.They came in today for appointment and the picc nurse was notified that the picc was seeping around the insertion site.Another nurse went to flush and it leaked right out so they ended up removing the picc and there is a hole the size 1 cm marking.The patient said she was infusing and then she had been wet so it seemed to be a steady leak.Additional information: there was no direct patient harm however the patient didn't properly receive her antibiotics for several days because after the hole in the line, the medication was coming back out around the insertion site of the picc.
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a leak was confirmed.The product returned for evaluation was one 4 fr single lumen powerpicc solo catheter.The investigation findings are consistent with damage accumulated through flexural fatigue.Flexural fatigue may occur due to cyclic kinking of the catheter tube in which physiological, placement, usage, and mechanical factors may gradually form a crack(s) in the catheter.The returned product sample was evaluated, and a split was observed at the 8 cm depth marking.The catheter split contained physical features associated with material fatigue, and the characteristics observed which supported this type of failure included: damage which was circumferentially aligned.Fracture edges which were rounded and polished due to repeated material wear.'c' shaped impressions leading into the fracture site which are consistent with material buckling due to movement which caused the fracture edges to be pressed together.Overall elliptical shape to the fracture cross-section (a result of repeated kinking of the tubing).Repetitive kinking of the indwelling catheter appears to have contributed to the observed leak; however, the specific circumstances surrounding how kinking developed into a fracture within the catheter tubing were ultimately unknown.An examination of the catheter structure revealed no potential damage/defect related to manufacture of the product.The damage location suggested that catheter securement, access, and maintenance techniques may have contributed.This complaint will be recorded for future trending and monitoring purposes.
|