| Model Number |
EGIA45AXT |
| Medical Device Problem Code |
Failure to Form Staple (2579)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Tissue Breakdown (2681)
|
| Date of Event |
02/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: egiaustnd, egiaustnd endogia ultra univ std stap (lot#p4k1164s); egiaustnd, egiaustnd endogia ultra univ std stap (lot#p4k1164s); egia45axt, egia45axt egia45 artic extra thicksulu (lot#n4f2071y); egia45axt, egia45axt egia45 artic extra thicksulu (lot#n4f2071y) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic low anterior resection (lar) procedure, when the first 45axt device was used to detach the rectum, the staple burst out and did not form a proper b-shape.This resulted in bleeding, and was sutured manually.Subsequent firings continued with two new 45axt devices, when firing, the handle could not be pressed.The first handle, which had been used when the previous staples burst out, was tried first, but the device still could not be fired after replacing it with a new handle.Both handles could not be pressed.After replacing the handle and reload, firing was successfully completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 h3.Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the staple burst out and did not form a proper b-shape.The reported issue was not confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|