• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA MECTACER BIOLOX DELTA FEMORAL BALL HEAD; BALL HEADS: MECTACER MECTACER HEAD BIOLOX DELTA DIA.28 12/14-M

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA MECTACER BIOLOX DELTA FEMORAL BALL HEAD; BALL HEADS: MECTACER MECTACER HEAD BIOLOX DELTA DIA.28 12/14-M Back to Search Results
Model Number 01.29.202
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Medical device site infection (4845)
Date of Event 02/05/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 11 february 2025.Lot 2317256: (b)(4) items manufactured and released on 06/07/2023.Expiration date: 18/06/2028.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Infection is a known possible complication of any surgery.Although no root cause can be established, there is no indication that any issue with the device may have caused or contributed to the event, and the document review does not indicate any potential manufacturing related cause.
 
Event or Problem Description
The patient came in due to signs of infection and the pathogen is unknown.About 1 year and 9 months after the primary surgery, the surgeon revised the competitor cup and liner and medacta head.The surgery was completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MECTACER BIOLOX DELTA FEMORAL BALL HEAD
Common Device Name
BALL HEADS: MECTACER MECTACER HEAD BIOLOX DELTA DIA.28 12/14-M
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key21523850
Report Number3005180920-2025-00134
Device Sequence Number8339926
Product Code LZO
UDI-Device Identifier07630030809132
UDI-Public07630030809132
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K112115
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number01.29.202
Device Lot Number2317256
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/05/2025
Initial Report FDA Received Date03/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/06/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-