| Model Number |
CAR-505 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
02/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No product was received for evaluation.A photo evaluation of the product confirmed the presence of a blood leak from the cartridge.A device history record (dhr) review was conducted for the reported lot number and confirmed the product was released for distribution having met quality and manufacturing specifications and requirements.The instructions for use states "check the system for blood and fluid leaks during treatment, and pay close attention to the blood line and access connections." all treatments must be administered under a physician's prescription and performed by a trained and qualified person, considered to be competent in the use of this device by the prescribing physician so that alarms and harmful conditions can be responded to promptly.
|
| |
|
Event or Problem Description
|
|
A report was received on (b)(6) 2025 from the critical care nurse of a patient of unspecified pathology, who stated blood was observed leaking from the cartridge during a continuous renal replacement therapy (crrt) on (b)(6)2025.Although requested, additional information was not provided.
|
| |
|