| Model Number |
M00583040 |
| Medical Device Problem Codes |
Break (1069); Failure to Conduct (1114); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a hydratome rx 44 was used in the duodenum during an endoscopic retrograde cholangiopancreatography (ercp) procedure.The exact procedure date was unknown.During the procedure, the device sparked and the cutting wire of the hydratome rx 44 broke.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b3: approximated based on the date the manufacturer became aware of the event.Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a hydratome rx 44 was used in the duodenum during an endoscopic retrograde cholangiopancreatography (ercp) procedure.The exact procedure date was unknown.During the procedure, the device sparked and the cutting wire of the hydratome rx 44 broke.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: approximated based on the date the manufacturer became aware of the event.H6: imdrf device code a0401 captures the reportable event of cutting wire broken.H11: (investigation result): the returned hydratome rx 44 was analyzed, and a visual and microscopic evaluation noted that the cutting wire was blackened, kinked, and broken from the proximal pierce hole.Additionally, the working length was torn from the end of the rx channel to the tip.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was blackened, kinked, and broken from the proximal pierce hole.Based on the condition of the device, the wire break could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wire integrity and cause a premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can lead to the broken section hitting the working channel of the scope.This can kink the cutting wire.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.The working length torn from the tip to the end of the rx channel is typically generated when the user pulls or removes the guidewire through the c-channel and the technique used during loading or removing the guidewire into the device.Based on all gathered information, the most probable root cause is adverse event related to procedure.
|
| |
|