| Model Number |
420207-10 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, the tenaculum forceps are faulty, the tip no longer responding to the robot after 2 uses.The procedure was competed with no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The tenaculum forceps instrument was analyzed and found to have a broken distal pitch cable at the clevis hub.The broken cable segment that contains the crimp was still installed within the clevis.The cable was fully broken.As a result of the full break, functional testing for motion related issues cannot be performed.There was no discoloration, corrosion, or contamination on the cable to suggest improper reprocessing as a contributing factor.Further inspection found no abnormally sharp or damaged components that could have contributed to the cable breaking.The complaint regarding the faulty tip was confirmed by failure analysis.
|
| |
|
Search Alerts/Recalls
|