• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER S.A.S. TORNIER FLEX SHOULDER SYS REV TRAY 0TH 1.5MM OFFSET; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER S.A.S. TORNIER FLEX SHOULDER SYS REV TRAY 0TH 1.5MM OFFSET; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number DWF510
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 02/07/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
Revision due to infection.
 
Event or Problem Description
Revision due to infection.
 
Additional Manufacturer Narrative
Correction - g1 reporting contact.The reported event could not be confirmed since the device was not returned for evaluation and no other evidence were provided.A device inspection was not possible since the affected device was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event must be available in order to determine the root cause of the complaint event.If any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORNIER FLEX SHOULDER SYS REV TRAY 0TH 1.5MM OFFSET
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR  38330
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
cécile lefeuvre
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key21529599
Report Number3000931034-2025-00100
Device Sequence Number15537490
Product Code KWS
UDI-Device Identifier03700386941054
UDI-Public(01)03700386941054(17)250602(21)5558AV035
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K190521
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date06/02/2025
Device Catalogue NumberDWF510
Device Lot Number5558AV
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/07/2025
Supplement Date Received by Manufacturer04/08/2025
Initial Report FDA Received Date03/05/2025
Supplement Report FDA Received Date05/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/02/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-